Specifications
-
Battery Type -
9 volt DC, 4.2 Ah, composed of disposable long-life lithium manganese dioxide primary cells
Capacity -
Minimum 200 shocks or 4 hours of operating time (EN60601-2-4:2003)
Install-by Date -
Battery is labeled with an install-by date of at least five years from date of manufacture
Standby Life -
Four years typical when battery is installed by the install-by date. (Will power the AED in standby state within the specified Standby Temperature range, assuming one battery insertion test and no defibrillation uses)
-
-
Product Specifications Kit Contents -
Defibrillator, battery (one, pre-installed), SMART Pads (one set, pre-installed), set-up/maintenance guide with expiration date tags, owner's manual, quick reference guide
Waveform -
SMART Truncated exponential biphasic. Waveform parameters adjusted as a function of patient impedance
Defibrillation peak current - Adult -
32 A (150 J nominal) into a 50 ohm load
Defibrillation peak current - Pediatric -
19 A (50 J nominal) into a 50 ohm load (using optional infant/child SMART Pads Cartridge)
Quick Shock -
Able to deliver a shock after the end of a CPR interval, typically in eight seconds
Shock-to-Shock Cycle Time -
Typically less than 20 seconds between shocks in a series
Voice Instructions -
Detailed voice messages guide responder through use of the defibrillator
CPR Guidance -
Instructions for adult and infant/child CPR available at user selection
Shock Delivery -
Via adhesive pads placed on patient's bare skin as illustrated on pads
Controls -
Green SMART Pads cartridge handle, green On/Off button, blue i-button, orange Shock button
Indicators -
Ready light, blue i-button, caution light
-
-
Environmental/Physical Requirements Standby Temperature -
0 - 43 °C (32 - 109 °F)
Operating Humidity -
0 to 95 % relative humidity (non-condensing)
Standby Humidity -
0 to 75 % relative humidity (non-condensing)
Standby Altitude -
> 48 hours from 0 to 8,500 ft and <48 hours from 8,500 to 15,000 ft
EMI (Radiated/Immunity) -
Meets EN55011 Group 1 Level B Class B and EN61000-4-3
-
-
SMART Pads Energy delivered -
Adult: nominal 150 joules into a 50 ohm load; Infant/Child: nominal 50 joules into a 50 ohm load
How Supplied -
Disposable cartridge, containing adhesive defibrillation pads, clicks into defibrillator for an integrated pads solution
Active Surface Area -
85 cm² (13.2 in²) each
Cable Length -
Adult pads: 137 cm (54"); Infant/Child pads: 102 cm (40")
Use-by Date -
Cartridge is labeled with a use-by date of at least two years from date of manufacture
Training pads (optional) -
Special pads put HeartStart into training mode and disable its energy delivery capability. Training pads feature 8 real-world training scripts. Use with training mat (included) or with adapters on manikins
-
-
Environmental/Physical Requirements Sealing -
Solid objects per EN60529 class IP2X; Drip-proof per EN60529 class IPX1
Operating Temperature -
0 - 50 °C (32 - 122 °F)
Operating Altitude -
0 to 4,500 m (0 to 15,000 ft)
Shock/Drop Abuse Tolerance -
Withstands 1 meter drop to any edge, corner, or surface
Vibration -
Meets EN1789 random and swept sine, road ambulance specification in operating and standby states
-
-
Automated and User-activated Self-tests Daily Automatic Self-tests -
Tests internal circuitry, waveform delivery system, pads cartridge, and battery capacity
Pads Integrity Test -
Specifically tests readiness-for-use of pads (gel moisture)
Battery Insertion Test -
Upon battery insertion, extensive automatic self-tests and user-interactive tests check device readiness
Status Indicator -
Blinking green "Ready" light indicates ready for use. Audible "chirp" indicates need for maintenance
-
-
Data Recording and Transmission Infrared -
Wireless transmission of event data to a PC or Palm PDA, using the IrDA protocol
Data Stored -
First 15 minutes of ECG and the entire incident's events and analysis decisions
-